Wide Range Resistance Meter Designed to Support ANSI/ESD S20.20 Program Standard
The PAS-853B is a wide range ohmmeter designed specifically for ESD Plant Auditors who must make many measurements quickly while supporting ANSI/ESD S20.20 Program Standard and periodic verification requirements. The PAS-853B performs in accordance with standard resistance measurement practices outlined in ANSI/ESD 541 Packaging and meets all audit requirements of ESD TR53.
Variable & Constant test voltages
- <10 Volts Variable from 0.01 to 9.9 x 103 ohms
- 10 Volts ± <0.5V (±<5% Constant Voltage) from 1.0 x 104 to 9.99 x 105 ohms
- 100 Volts ± <5V (±<5% Constant Voltage) from 1.0 x 106 to 9.99 x 1012 ohms
Features:
- Wide Resistance Measurement Range from 0.01 ohm to 9.99 x 1012 ohms
- Nominal Full Range Tolerance Averages <±5%
- Fully Automatic Resistance Range, Test Voltage and Electrification Period control
- Automatic or Continuous Electrification Control
- Rugged High Performance 80 inches (200cm) Test Leads
- Powered by one long lasting 9V alkaline battery
- Large, easy to read LCD Display
- Meets requirements of ESD Association Standards for material and audit resistance measurement requirements
- Certificate of Calibration Traceable to NIST included
Technical Data
PAS-853B | |
Controls | ON/OFF Battery Bus Cut-off Slide Swtch ON SIDE OF CASE ON: Energizes instrument and LCD displays Software Version then GO A with AUTO indification (Default) OFF: Isolates battery and DC voltage from instrument controls and circuit. Use for transport and storage. |
TEST Pad | With Test Leads installed: Press once & Release. Measurements performed automatically In conjunction with MODE changes from Auto operation to Continuous |
MODE Pad |
Steps through Auto/Voltage ranges with each press of MODE pad:
- Resistance Range : 0.01E-1 to 9.99e9Ω
- Resistance Range: 0.98E4 to 9.99E10Ω
- Resistance Range: 0.98E6 to 9.99E12Ω |
Connections | Positive (+): Test Voltage (RED) - Supplies test voltage to electrodes/fixtures Negative (-): Current sensing (BLACK) - Conducts current from electrode, fixtures or Ground to instrument Chassis (): Shielding connection to chassis (WHITE) - Reduces transmitted energy and effect on current measurement Note: ¾ inch spacing between RED and BLACK accommodates dual banana BNC adapters |
Indication |
Colored LED's indicate instrument operational MODE and Test Voltage (YELLOW) used for the current measurement,
|
Typical Tolerances | 1Nominal performance by range: 0.01E- 1 -- 1.0E 0 Ω <±10%,±2 Counts 1.0E 0 - 9.99E10 Ω <±2%,±2 Counts 1.0E11 -- 9.99E12 Ω <±10%,±4 Counts |
Test Voltages | Automatically selected and controlled in AUTO mode. May be manually selected by pressing MODE pad in the following order: <10V Variable based on resistance load - Provides 0.001 to ≤10V - Functional in AUTO within range 0.01Ω - <1.0E+4 Ω - Manual Operation Resistance Range: 0.01E-1 to 9.99@+9 Ω 10V Constant Voltage (Under Load) - Test Voltage Varies ± <5% in AUTO or Manual Operations - Functional in AUTO within range 1.0E+4Ω - <1.0E+6 Ω - Manual Operation Resistance Range: 0.9E+4 to 9.99E+10 Ω 100V Constant Voltage (Under Load) - Test Voltage Varies ± <2% in AUTO or Manual Operations - Functional in AUTO within range 1.0E+6 Ω - <1.0E+13 Ω - Manual Operation Resistance Range: 0.90E6 to 9.99E12 Ω Returns to Auto Mode |
Modes of Operations |
GO A: Performs all automatic and manual voltage select functions, stops and displays measurements after Electrification Period (EP) is complete, then displays data with Green LED ON for 10 seconds. GO C: Performs all automatic and manual voltage select functions, continuously displays measurements until the operator presses MODE or TEST pads. GO C disables electrification period controls. |
Power | One 9V alkaline transistor battery; battery life > 50 hours |
Dimensions | 7.25 in. (184 mm) x 4.3 in, (109 mm) x 1.4 in. (36 mm) |
Weight | 13.2 oz. (375 gram) with battery installed |
Notes |
1Tolerance and accuracy measurements were conducted under controlled conditions with close tolerance reference resistors.
|